Digital 32-Channel EEG Machine Mapping System Recorder+Software,Tripod,KT88-3200Model:KT88-3200
Manufacture:CONTEC Medical Systems
Introduction
KT88-3200 Digital Brain Electric Activity Mapping is an equipment checking 32-lead EEG, it collects EEG signal with electrodes, via integrated amplification, A/D transformation, PC auto-analysis, FFT, to form electroencephalogram which displays with color depth. The product is applicable for checking such diseases as epilepsy, intracranial inflammation, cerebrovascular diseases and brain tumors.
Performance Parameter
◆ Numbers of channels:32 channels of EEG
◆ Speed of sampling:200 Hz/sec.
◆ Precision of sampling:12 bit
◆ Input impedance: ≥10MΩ
◆ Patient leak current:< 10uA
◆ Level of noise: ≤5uV Vp-p
◆ CMRR:≥90dB
◆ Magnifying multiple:10 000
◆ Filtering constant:all digital and free enactment
◆ Speed of displaying(paper speed):5,10,15,30,60,120mm/s
◆ Displaying gain:1,2,5,10,12,15,20 mm/50uv
◆ Speed of review:1, 2, 3, 10, 20, 40, 60
◆ 50Hz interference suppression ratio: ≥30dB
◆ Safety type:class II, type BF applied part
Function Features
◆ Standard 32 leads of EEG. Method of 10/20 electrodes placement under international standard system, can change leads ways while real-time playing or replaying. Supporting sorts of lead mode, such as Al, A2, A1+A2, A1A2, BN, CZ, user-defined, and so on.
◆ Using bioelectricity amplifier brainwave, consecutive record time can be up to 24 hours, integrated automatic calibration system.
◆ Powerful playback function: It is adjustable for the amplitude and replaying rate. The special subdividing time line divides the waveform in one second into 5 parts and easy for the doctors to look over the waveform.
◆ Multifunctional digital filter system can be set freely and provides many filtering methods.
◆ EEG signal clipping function,clipping any a section of EEG wave to analyse and store, clipping multi-section waves to perform multi-parameter analyse and pick-up automatically.
◆ Electronic frequency ruler, convenient to measure the basic information of any appointed EEG waves. Sectional enlargine window, which can be adjusted by yourself, is provided to increase the accuracy of EEG period /amplitude /frequency measurement.
◆ Mark event (openning eyes, closing eyes and flashing) EEG with different colors, and name the user-defined event. Color of background and waveform occurred by different event is adjustable. It is able to depend on the event's name to locate speedily the waveform of the corresponding time.
◆ Powerful automatic analyse function, can carry through the power spectrum analyse and pathologic wave detection for appointed waveform. Sorts of graph can be displayed in the same screen, including kinds of BEAM, numerical BEAM, compressed spectrum graph, trend graph, and so on.
◆ Professional isolation transformer, dual power supply isolation system and optoelectronic data transmission to ensure security. Use USB interface to transmit data which just need to be inserted. Distance of digital signal transmission is up to a maximum of 30 meters. Patients and doctors can be operated in different rooms.
◆ Multifunctional flashlight of USB interface, and frequency can be controled manually or automatically. A flashlight stimulus scheme can be set and performed in the process of collection.
◆ Perfect case management, multi-method of information query and quick statistics means, convenient case input and output function, supporting MO or CD-RW store, which is easy for data research.
◆ The report of picture and character integration, can be edited in mode and changed into Word.
◆ The case can be transfromed into the EDF data format, which is convenient for the data interchange, academic exchanges and further analyse.
◆ System setup function, multi-function, such as display mode, system parameters, etc., can be set to meet the users' special need.
◆ Mark positioning function: add the mark and note for the waveform of appointed time, and locate speedily the waveform at the mark time.
◆ Optional video function: USB CCD camera is easy to install, convenient to use and exact to record. With agile playback function, it can be used to replay the waveform of any time along with the corresponding isochronous collected image.
◆ SpO2 function is optional.
Accessories:
Standard accessories:
◆Forty EEG leads
◆Two data lines
◆Forty EEG electrodes
◆Two headgears
◆One flashlight
◆One flashlight power supply
◆One software disk
◆One User Manual
◆One earth wire
◆Two brackets
Optional accessories:
◆SpO2 module (with one CMS-P SpO2 probe and one SpO2 software disk)
◆Video module (with one softdog, one camera and one software disk)
Qualification Certificate
Internal registration
Buy safe products
The following FDA Disclaimer is required for all eBay listing in Healthcare category and is included for REFERENCE: The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item.
The Fingertip Pulse Oximeter is registered on the Australian Register of Therapeutic Goods (ARTG)with the code 197923, and certified by FDA of United States and CE,TUV of Europe. The Fingertip Pulse Oximeter that is FDA 510K Approved
Shipping:
1). We will ship within 1 business day after your payement ;
ship from China warehouse by China Post 2-4 weeks Arrival.
2).We don't add taxes, VAT or other hidden charges. You pay us what you see on the invoice, i.e. goods subtotal + shipping cost (not include duties). Please find out as much as you can about your import taxes in your own country before bidding the item. You need to pay for the import duties in case that occurred for certain goods, or you may kindly give us some suggestions about that, we promise to do our best to help you!
About Warranty:
We guarantee new equipment other than accessories to be free from defects in workmanship and materials for a period of twelve months (six months for spare parts) from the date of shipment under normal use and service. Our company's obligation under this warranty is limited to repairing, at our company's option, any part which upon our company's examination proves defective.
Return Policy:
If you receive a defective item which you want to return, please contact us within 5 days from the day you receive the item.
All return items must be returned with its original packaging and accessories. If we confirm the damage happened during the shipment, we will bear the shipping cost for return.
We will refund the money to you when we get the return items. Or replace item for you free of charge.
Only English User manual,Please contact us if you need other language version.